No Template Control Shows Ct 38: Do I Need to Rerun?
Probably not — but it depends on how far that Ct 38 sits from your lowest-abundance sample. If your gene of interest has Ct values in the 20–28 range, a NTC at Ct 38 represents roughly a 1,000- to 10,000-fold lower signal. That's almost certainly primer-dimer or instrument noise, not meaningful contamination. Your data is fine.
If your samples are sitting at Ct 32–35, though, a NTC at 38 starts to close that gap uncomfortably. A difference of only 3–5 Ct between your weakest sample and the NTC means the "contamination" signal is within 10- to 30-fold of your real signal — and that can distort your quantification. In that case, yes, you should investigate and likely rerun.
Why NTCs Rarely Come Back Perfectly Clean
A pristine NTC — flat amplification curve, no Ct assigned — is the ideal. In practice, especially with SYBR Green or other intercalating dye chemistries, late-cycle fluorescence blips are common. There are a few reasons:
- Primer-dimer formation. Even well-designed primers can form short self-annealing products after 35+ cycles when there's no competing template. These are typically 40–80 bp, well below your amplicon size, and show up as a distinct low-temperature peak on your melt curve (more on that below).
- Baseline drift. Instruments like the CFX96 and QuantStudio 3/5 sometimes call a Ct in the high 30s when the raw fluorescence barely crosses threshold. This is an artifact of baseline subtraction, not real amplification.
- Low-level environmental DNA. If you're amplifying ubiquitous targets — 18S rRNA, GAPDH, ACTB — trace human DNA from skin cells, aerosols, or reagent manufacturing can produce a legitimate but minuscule amplification signal. This is a known issue with 18S in particular.
- Probe hydrolysis artifacts (TaqMan). Less common, but TaqMan probes can degrade over freeze-thaw cycles, releasing free fluorophore. This causes a slow upward drift that the software may interpret as amplification. The curve shape is usually obviously wrong — linear rather than sigmoidal.
None of these are reasons to panic. They're reasons to look more carefully before deciding.
The Melt Curve Is Your Best Diagnostic
If you're running SYBR Green (PowerUp SYBR, Luna Universal, iTaq, etc.), the melt curve tells you almost everything you need to know about a NTC signal at Ct 38.
Check the melt peak of your NTC against your positive samples. You're looking for one of three scenarios:
NTC melt peak is at a distinctly lower Tm than your samples (e.g., NTC peak at 72°C, samples at 82°C). This is primer-dimer. Ignore it. Your data is clean.
NTC melt peak matches your samples' Tm exactly. This means genuine amplification of your target sequence occurred in the NTC well. This is real contamination — likely from template carryover or aerosol. The question is whether it matters quantitatively (see the next section).
NTC shows no discernible melt peak, just noise. The Ct call was an artifact. Ignore it.
If you're using TaqMan probes, you don't have melt curves. Instead, look at the amplification curve shape. A real amplification event produces a sigmoidal curve with a clear exponential phase. A NTC artifact at Ct 38 on TaqMan usually looks like a slow linear creep — not sigmoidal. Most software will still assign a Ct, but you can visually dismiss it.
The "10 Ct Rule" and When to Apply It
A widely used rule of thumb: if your NTC Ct is at least 5–10 cycles higher than your most dilute sample, the contamination is negligible. Here's the math behind it.
Each Ct difference of ~3.3 (assuming 100% efficiency) represents a 10-fold difference in starting template. So:
| Ct gap (sample vs. NTC) | Fold difference | Contamination as % of signal |
|---|---|---|
| 5 Ct | ~32-fold | ~3% |
| 7 Ct | ~128-fold | ~0.8% |
| 10 Ct | ~1,024-fold | ~0.1% |
If your lowest sample is Ct 28 and your NTC is Ct 38, that's a 10 Ct gap — the contaminating signal contributes about 0.1% to your measurement. That's noise within your pipetting error.
If your lowest sample is Ct 33 and the NTC is Ct 38, that's a 5 Ct gap. Now contamination accounts for roughly 3% of signal. For most gene expression studies using the ΔΔCt method (Livak and Schmittgen, 2001), a 3% contribution translates to a shift of about 0.04 Ct — still well within acceptable replicate variation (< 0.5 Ct SD).
The real danger zone is a gap of 3 Ct or less. At that point, contamination contributes 10%+ of the signal, which can shift your fold-change calculations meaningfully. If you're in that territory, rerun with fresh reagents and freshly diluted primers.
Systematic NTC Signals: When the Problem Is Bigger
A one-off NTC blip at Ct 38 is one thing. If you're seeing consistent NTC amplification across multiple plates or multiple primer pairs, you have a workflow problem. Common culprits, in order of likelihood:
Contaminated primer stocks. If you resuspended lyophilized primers with non-nuclease-free water, or if the tube has been opened dozens of times, your primer working stocks may carry template DNA. Make a fresh 10 µM working stock from the 100 µM master and test again.
Contaminated master mix aliquot. Especially if you're using a communal bottle of PowerUp SYBR or Luna Universal that lives in a shared freezer. Aliquot into single-use volumes and discard any tube that's been opened more than 5–8 times.
PCR product contamination (the amplicon problem). If you run gel electrophoresis on PCR products at the same bench where you set up reactions, aerosolized amplicons will find their way into everything. This is the most insidious source because it produces NTC signals with perfect melt curves — indistinguishable from real template. Separate your pre-PCR and post-PCR workspaces physically. Use dedicated pipettes.
Plasmid or gBlock contamination. If you've been handling standard curve plasmids or synthetic gene fragments in the same area, even femtograms of those molecules can template a reaction. These are the worst offenders because they're at astronomically high copy numbers per mass.
If you suspect amplicon or plasmid contamination, switching to a dUTP-containing master mix with uracil-DNA glycosylase (UDG) treatment can help. PowerUp SYBR Green Master Mix includes this by default. UDG degrades any carryover amplicons containing dUTP before cycling begins.
What to Report in Your Paper
Reviewers occasionally ask about NTC results. The standard practice is simple: state that NTCs were included on every plate and showed no amplification, or if they showed late-cycle signal, state the Ct and note that it was >X cycles beyond the most dilute sample.
Something like: "No-template controls were included for all primer pairs on each plate. NTCs showed no amplification or exhibited Ct values > 37, at least 8 cycles beyond the lowest-abundance experimental sample."
That's sufficient. No reviewer is going to reject a paper because your HPRT1 NTC flickered at Ct 39.
A Quick Decision Framework
When you see a NTC with a Ct value, run through this:
- What's the Ct gap between NTC and your lowest real sample? If ≥ 7 Ct, you're almost certainly fine.
- What does the melt curve show? If it's primer-dimer (lower Tm), ignore it. If it matches your target, proceed to step 3.
- Is it reproducible? If only 1 of 3 NTC replicates shows signal, it's likely a stochastic event — a single molecule of contamination partitioning into one well. If all 3 NTC replicates amplify at similar Ct values, you have a systematic problem.
- Does it affect your conclusions? If you're comparing a treated group (Ct 22) vs. control (Ct 24) and the NTC is at 38, the contamination is irrelevant to your biology. If you're trying to detect a low-abundance transcript at Ct 35 and the NTC is at 37, your assay isn't giving you reliable data for that target regardless.
If you're running ΔΔCt analysis and want to check whether a borderline NTC is affecting your fold-change calculations, upload your data to VoilaPCR — it flags NTC issues automatically and shows you exactly how much (or how little) they shift your results.
Bottom line: Ct 38 in your NTC is almost never a reason to rerun. Check the melt curve, check the gap, and move on.